Description

【Principle of the method】
Anti-human C3 antibodies when mixed with samples containing C3, form insoluble complexes. These complexes cause an absorbance change, dependent upon the C3 concentration of the patient sample, that can be quantified by comparison from a calibrator of know C3 concentration.
【Reagent composition】
Diluent(R1) :PBS buffer *0 mmol/L, PEG ***0,Surfactant.
Antibody(R2):PBS buffer *0 mmol/L, Goat serum, anti-human C3,Preservative.
【Storage and Shelf Life】
1.Sored tightly closed at **8ºC.
2.Non-open kit stored tightly closed at  **8ºC has *2 month shelf  life, once opening, the kit has one more month at**8ºC storage。
【Applied analyzer】
Spectrophotometer or biochemical analyzer with **0nm wavelength.
【Sample】
Fresh serum or plasma. Do not use highly hemolized or lipemic samples.If not determine in time, store at**0℃. Avoid repeat freezing and melting.
【Procedure】
 

  1. Assay method:two point,wavelength**0nm
2.   Application on chemistry analyzer:
 
Sample Reagent(R1/R2) Temperature Reaction time(T1/T2)
3μL **0μL/**0μL *7℃ **0s./ **0s.
 
 
Unit Reaction Direction Standard curve simulation formula
g/L Positive Multi point calibration, non-linearity calibration mode
3.Determination:Add3uLsamples into**0μL R1,mix thoroughly and incubate the cuvette  at *7℃for **0 seconds. Read the absorbance A1against the blank.. Then mix**0μL R2thoroughly and incubate the cuvette for **0 seconds, read the absorbance A2. The absorbance of complex is△A(A*-A1).
Note: different chemistry analyzer has own specifications, but should be subject to the basic specifications herein above, and  the kit could be used after setting on the chemistry analyzer.
【Reference values】
Between 0.*9g/L~1.*2g/L . Each laboratory should establish its own reference range.
 